BINDER FP 260 drying and heating oven
The drying and heating oven impresses above all with its precise temperature control, short heat-up times and versatile programme functions. Users in laboratories, research facilities and industrial quality assurance benefit from its high reliability and uncomplicated operation. This oven offers major advantages over comparable appliances, particularly when testing materials, sterilising or for precise drying processes, as its sophisticated features create optimum conditions for controlled thermal conditions.
High process reliability for demanding applications
The device is specially designed to precisely control material and ageing processes and thus ensure reproducible test results. The APT.line process creates a uniform temperature distribution, which ensures consistent results even when fully loaded. Thanks to the user-friendly controller, temperature profiles can be created effortlessly - both as a time segment and as a real-time programme. This enables clear, task-specific controls that are appreciated in research laboratories, quality assurance departments or production environments.
Most important features and technical data at a glance:
- Temperature range: +10 °C above ambient to +300 °C
- Very low temperature deviation over time (±0.4 K at 150 °C)
- Fast heat-up time to 150 °C in approx. 19 minutes
- Recovery time after door open for 30 seconds at 150 °C: 6 minutes
- Adjustable fan speed to regulate circulating air individually
- Computer interface (Ethernet) for comprehensive data recording and analysis
- Temperature selection limiter class 2 and temperature selection monitor class 3.1 (DIN 12880)
In addition, the appliance has an electromechanical exhaust air flap control, which ensures constant humidity and a defined removal of moisture in many drying and testing applications. The internal data logger stores measured values in an open read-out format so that results can also be quickly and easily analysed externally. With two chrome-plated slide-in grids as standard and a maximum total load of up to 270 kg, a wide variety of samples and materials can be placed flexibly.
The high quality and durability is also emphasised by the ergonomic handle and robust outer housing - a reliable choice for many years of use under demanding conditions. The relevant standards include DIN 12880, according to which the temperature selector is designed. Various alarm signals (audible-visual) can be selected directly on the controller to safely monitor critical temperature ranges.
Application examples in the laboratory and industry
- Material testing for plastics, metals or ceramic materials to analyse ageing processes at higher temperatures
- Sterilisation processes in microbiological or medical laboratories where temperatures of up to 300 °C are required
- Drying of complex components in electronics production to remove moisture and solvent residues.
- Heat treatment of samples before and after chemical analyses in research facilities (e.g. for experiments with polymers or organic compounds).
Frequently asked questions
How can temperature documentation be automated?
Measured values can be automatically recorded via the Ethernet interface and the internal data logger and read out for further analysis.
What to look out for when loading
The maximum total load of 270 kg and the 40 kg permitted per rack must be taken into account so that air circulation is not impaired and temperature maintenance is guaranteed.
Is it possible to precisely set the air exchange rate?
Yes, thanks to adjustable air circulation and precise fan speed adjustment, the air exchange rate can be adapted to the respective process. At 100 °C, the appliance achieves an air exchange rate of around 9 x-h.
What safety functions exist in the event of overheating?
The appliance has a class 2 temperature selector limiter and a class 3.1 temperature selector monitor (in accordance with DIN 12880), which issue an alarm if critical parameters are reached or exceeded and protect against damage.
Is it possible to stack several appliances?
Models with an interior volume of up to 115 litres can be stacked. For larger appliances such as this model, we definitely recommend checking the space individually to ensure optimum ventilation and compliance with the minimum clearances.
Conclusion
This material test chamber combines precise temperature control with a high level of reliability, which is essential in both laboratory and production environments. If you are looking for consistent results, rapid heating phases and flexible programming, this is a powerful and reliable tool. The modern controller concept and intelligent air circulation ensure constant temperature conditions with minimal deviations. Thanks to its versatile application options, robust design and easy operating processes, this oven is ideally suited to cover even complex drying or heating tasks safely and reproducibly.
| Load max.: | 251 - 300 kg |
| Width outside (incubators,...): | 810 mm |
| Width inside (incubators,...): | 650 mm |
| Maximum temperature (heat): | + 300 °C |
| Height outside (incubators,...): | 962 mm |
| Height inside (incubators,...): | 780 mm |
| Interior volume: | 201 - 300 l |
| Maximum power: | 2.5 kW |
| Suitable for: | Lacquer dryingDrying - moisture extraction |
| Cabinet type: | heating ovenHeating cabinet |
| Safety standard: | DIN 12880 |
| Depth outside (incubators,...): | 786 mm |
| Depth inside (incubators,...): | 512 mm |
| Shipping weight: | 82,00 Kg |
| Product weight: | 82,00 Kg |
| Dimensions: | × × cm |